Knowledge

Medical thermometer

Source 📝

645:
the most common type of thermistor. NTCs have temperatures that vary inversely with their resistance, so that when the temperature increases, the resistance decreases, and vice versa. NTCs are constructed from oxides of materials such as nickel, copper, and iron. • Positive temperature coefficient (PTC) devices are used in electric current control. They function in an opposite manner than NTC in that the resistance increases as temperature increases. PTCs are constructed from thermally sensitive silicons or polycrystalline ceramic materials. • There are several advantage and disadvantages to using an NTC thermistor thermometer. • Advantages include their small size and high degree of stability. NTCs are also long lasting and very accurate. • Disadvantages include their non-linearity, and unsuitability for use in extreme temperatures
700:
resolution (0.05 °C or 0.1 °F is sufficient) may be suitable for monitoring basal body temperatures; the specification should be checked to ensure absolute accuracy, and thermometers (like most digital instruments) should be calibrated at specified intervals. If only the variation of basal temperature is required, absolute accuracy is not so important so long as the readings do not have large variability (e.g., if real temperature varies from 37.00 °C to 37.28 °C, a thermometer which inaccurately but consistently reads a change from 37.17 °C to 37.45 °C will indicate the magnitude of the change). Some digital thermometers are marketed as "basal thermometers" and have extra features such as a larger display, expanded memory functions, or beeping to confirm the thermometer is placed properly.
613:, because they display numeric values) have been used. Many display readings with a resolution of only 0.1 °C (.2 °F), but this should not be taken as a guarantee of accuracy: the specified accuracy must be checked in the documentation and maintained by periodic recalibration. A typical inexpensive electronic ear thermometer for home use has a displayed resolution of 0.1 °C, but a stated accuracy within ±0.2 °C (±0.35 °F) when new. The first electronic clinical thermometer, invented in 1954, used a flexible probe that contained a Carboloy thermistor. 511:—typically five minutes in the mouth and ten minutes under the armpit. Maximum-reading is achieved by means of a constriction in the neck close to the bulb. As the temperature of the bulb rises, the liquid expands up the tube through the constriction. When the temperature falls, the column of liquid breaks at the constriction and cannot return to the bulb, thus remaining stationary in the tube. After reading the value, the thermometer must be reset by repeatedly swinging it sharply to shake the liquid back through the constriction. 29: 372: 636:
the temperature difference between the ends. •Advantages of thermocouples include their high accuracy and reliable operation over an extremely wide range of temperatures. They are also well-suited for making automated measurements both inexpensive and durable. •Disadvantages include errors caused by their use over an extended period of time, and that two temperatures are required to make measurements. Thermocouple materials are subject to corrosion, which can affect the thermoelectric voltage
602: 2114: 578: 627:
Platinum is the most commonly used material because it is nearly linear over a wide range of temperatures, is very accurate, and has a fast response time. RTDs can also be made of copper or nickel. Advantages of RTDs include their stable output for long periods of time. They are also easy to calibrate and provide very accurate readings. Disadvantages include a smaller overall temperature range, higher initial cost, and a less rugged design
273:(1700–1741). De Haen made particular strides in medicine with the thermometer. By observing the correlation in a patient's change in temperature and the physical symptoms of the illness, he concluded that a record of one's temperature could inform the doctor of a patient's health. However, his proposals were not met with enthusiasm by his peers and the medical thermometer remained a scarcely used instrument in medicine. 727: 357:) temperature is measured by holding the thermometer tightly under the armpit. One needs to hold the thermometer for several minutes to get an accurate measurement. The axillary temperature plus 1 °C is a good guide to the rectal temperature in patients older than 1 month. The accuracy from the axilla is known to be inferior to the rectal temperature. 320:
readings of the same value. When a temperature is quoted the location should also be specified. If a temperature is stated without qualification (e.g., typical body temperature) it is usually assumed to be sub-lingual. The differences between core temperature and measurements at different locations, known as
533:; the vigorous swinging needed to "reset" a mercury maximum thermometer makes it easy to accidentally break it and release poisonous mercury vapors. Mercury thermometers have largely been replaced by electronic digital thermometers, or, more rarely, thermometers based on liquids other than mercury (such as 699:
Glass oral thermometers typically have markings every 0.1 °C or 0.2 °F. Basal temperature is stable enough to require accuracy of at least 0.05 °C or 0.1 °F, so special glass basal thermometers are distinct from glass oral thermometers. Digital thermometers which have sufficient
675:
According to a 2001 research, electronic thermometers on the market significantly underestimate higher temperatures and overestimate lower temperatures. The researchers conclude that "the current generation of electronic, digital clinical thermometers, in general, may not be sufficiently accurate or
525:
The tube must be very narrow to minimise the amount of mercury in it—the temperature of the tube is not controlled, so it must contain very much less mercury than the bulb to minimise the effect of the temperature of the tube—and this makes the reading rather difficult as the narrow mercury column is
644:
Thermistor elements are the most sensitive temperature sensors available. A thermistor is a semiconductor device with an electrical resistance that is proportional to temperature. There are two types of products. •Negative temperature coefficient (NTC) devices are used in temperature sensing and are
666:
emitted from the location. Although these are not in direct contact with the area being measured, they may still contact part of the body (a thermometer which senses the temperature of the eardrum without touching it is inserted into the ear canal). To eliminate the risk of patient cross-infection,
423:
This tympanic thermometer has a projection (protected by a one-time hygienic sheath) that contains the infrared probe; the projection is gently placed in the ear canal and a button pressed; the temperature is read and displayed within about a second. These thermometers are used both in the home and
635:
Thermocouples are accurate, highly sensitive to small temperature changes, and quickly respond to changes to the environment. They consist of a pair of dissimilar metal wires joined at one end. The metal pair generates a net thermoelectric voltage between their opening and according to the size of
626:
RTDs are wire windings or other thin film serpentines that exhibit changes in resistance with changes in temperature. They measure temperature using the positive temperature coefficient of electrical resistance of metals. The hotter they become, the higher the value of their electrical resistance.
444:
thermometers, which use the infrared principle report temperature, are becoming increasingly common in clinical practice because of their ease of use and minimal invasiveness. Because of the variability of technique and environmental considerations, measurements by temporal artery thermometers may
653:
Some electronic thermometers may work by contact (the electronic sensor is placed in the location where temperature is to be measured, and left long enough to reach equilibrium). These typically reach equilibrium faster than mercury thermometers; the thermometer may beep when equilibrium has been
319:
The temperature can be measured in various locations on the body which maintain a fairly stable temperature (mainly oral, axillary, rectal, tympanic, or temporal). The normal temperature varies slightly with the location; an oral reading of 37 °C does not correspond to rectal, temporal, etc.
568:
Phase-change thermometers use samples of inert chemicals which melt at progressively higher temperatures from 35.5 °C to 40.5 °C in steps of 0.1 °C. They are mounted as small dots in a matrix on a thin plastic spatula with a protective transparent cover. This is placed under the
427:
There are factors that make readings of this thermometer to some extent unreliable, for example faulty placement in the external ear canal by the operator, and wax blocking the canal. Such error-producing factors usually cause readings to be below the true value, so that a fever can fail to be
341:
Oral temperature may only be taken from a patient who is capable of holding the thermometer securely under the tongue, which generally excludes small children or people who are unconscious or overcome by coughing, weakness, or vomiting. (This is less of a problem with fast-reacting digital
142:
is the first known individual to have put a measurable scale on the thermoscope and wrote of it in 1625, though he possibly invented one as early as 1612. His models were bulky, impractical and took a fair amount of time to take an accurate oral reading of the patient's temperature.
391:
noted "what for many remains a humiliating procedure ... insertion of a rectal thermometer. Also, if not taken the correct way, rectal temperature-taking can be uncomfortable and in some cases painful for the patient. Rectal temperature-taking is considered the method of choice for
569:
patient's tongue. After a short time the spatula is removed and it can be seen which dots have melted and which have not: the temperature is taken as the melting temperature of the last dot to melt. These are cheap disposable devices and avoid the need for sterilizing for re-use.
342:
thermometers, but is certainly an issue with mercury thermometers, which take several minutes to stabilise their reading.) If the patient has drunk a hot or cold liquid beforehand time must be allowed for the mouth temperature to return to its normal value.
276:
Thermometers remained cumbersome to transport and use. By the mid 19th century, the medical thermometer was still a foot long (30.28 cm) and took as long as twenty minutes to take an accurate temperature reading. Between 1866 and 1867,
491:
The traditional thermometer is a glass tube with a bulb at one end containing a liquid which expands in a uniform manner with temperature. The tube itself is narrow (capillary) and has calibration markings along it. The liquid is often
477:
or similar technology; at a given temperature the markings (numerals indicating the temperature) in one region are at the right temperature to become visible. This type may give an indication of fever, but is not considered accurate.
714:
A smart thermometer is able to transmit its readings so that they can be collected, stored and analysed. Wearable thermometers can provide continuous measurement but it is difficult to measure core body temperature in this way.
556:. Gallium is considered non-toxic and when disposed of does not face environmental concerns. Like mercury, gallium is a liquid at body temperature (melting point 29.7 °C), but according to the manufacturer an 332:
of the differences). For example, one study found that the clinical bias of rectal temperatures was greater than for ear temperature measured by a selection of thermometers under test, but variability was less.
160:(1686–1736), a Polish-born Dutch physicist, engineer, and glass blower, made contributions to thermometers as well. He created an alcohol thermometer in 1709 and later innovated the mercury thermometer in 1714. 1471:"A prospective observational study testing liquid crystal phase change type thermometer placed on skin against oesophageal/pharyngeal placed thermometers in participants undergoing general anesthesia" 1313:
Kiekkas, P; Stefanopoulos, N; Bakalis, N; Kefaliakos, A; Karanikolas, M (April 2016). "Agreement of infrared temporal artery thermometry with other thermometry methods in adults: systematic review".
522:
have been considered the most accurate liquid-filled types. However, mercury is a toxic heavy metal, and mercury has only been used in clinical thermometers if protected from breakage of the tube.
404:
The ear thermometer was invented by Dr. Theodor H. Benzinger in 1964. At the time, he was seeking a way to get a reading as close to the brain's temperature as possible, since the
1077:
Rotello, LC; Crawford, L; Terndrup, TE (1996). "Comparison of infrared ear thermometer derived and equilibrated rectal temperatures in estimating pulmonary artery temperatures".
696:
is much less affected than daytime temperature by environmental factors such as exercise and food intake. This allows small changes in body temperature to be detected
1597: 281:(1836–1925) designed a medical thermometer that was much more portable, measuring only six inches long and taking only five minutes to record a patient's temperature. 529:
It has been decided by many states to prohibit the use and sale of mercury thermometers due to the risk of handling and spilling, and the potential to cause
416:, which are shared with the hypothalamus. Before the ear thermometer's invention, easy temperature readings could only be taken from the mouth, rectum, or 151: 292:. With his findings, he was able to conclude a healthy human's temperature fell within the range of 36.3 to 37.5 °C (97.34 to 99.5 °F). 1520: 507:
To use the thermometer, the bulb is placed in the location where the temperature is to be measured and left long enough to be certain to reach
856: 461:
of 97–100% for detecting fever and hypothermia. Because of this, it is suggested that they should not be used in acute care settings like the
2073: 1049: 465:, or in patients with a high suspicion of temperature imbalance. Evidence supports higher accuracy and precision amongst pediatric patients. 549:
One company markets a thermometer of this sort which it bills as "the first analogue thermometer without mercury," as it instead uses the
379:
Rectal thermometer temperature-taking, especially if performed by a person other than the patient, should be facilitated with the use of a
420:. Previously, if doctors wanted to record an accurate brain temperature, electrodes needed to be attached to the patient's hypothalamus. 132:
circa 1592–1593. It lacked an accurate scale with which to measure temperature and could be affected by changes in atmospheric pressure.
609:
Since compact and inexpensive methods of measuring and displaying temperature became available, electronic thermometers (often called
345:
The typical range of a sub-lingual thermometer for use in humans is from about 35 °C to 42 °C or 90 °F to 110 °F.
473:
The thermometer is applied to the patient's brow. It is typically a band coated with different temperature-sensitive markings using
1713: 288:
published his studies that consisted of over one million readings from twenty-five thousand patients' temperatures, taken in the
1196:
Zengeya, S.T.; Blumenthal, I. (December 1996). "Modern electronic and chemical thermometers used in the axilla are inaccurate".
778: 1998: 1939: 1653: 937: 1896: 2140: 1901: 2080: 177: 328:. Measurements are subject to both site-dependent clinical bias and variability between a series of measurements ( 2100: 2068: 1448: 285: 1289: 1809: 1161:
Shann, Frank; Mackenzie, Angela (1 January 1996). "Comparison of Rectal, Axillary, and Forehead Temperatures".
380: 1349: 383:. Although rectal temperature is the most accurate, this method may be considered unpleasant or embarrassing 2046: 2031: 1769: 1571:
Latman, NS; Hans, P; Nicholson, L; Delee Zint, S; Lewis, K; Shirey, A (2001). "Evaluation & Technology".
845: 667:
disposable probe covers and single-use clinical thermometers of all types are used in clinics and hospitals.
325: 308: 2021: 519: 172:, having recorded the system in 1724. The scale is still only mainly used for everyday applications in the 1627: 1623: 2150: 2051: 1247: 441: 295:
Dr. Theodor H. Benzinger (13 April 1905 - 26 October 1999) invented the ear thermometer in 1964. Born in
1287: 745: 586: 504:
is often used, which indicates the maximum temperature reached even after it is removed from the body.
408:
at the brain's base regulates the core body temperature. He accomplished this by using the ear canal's
270: 944:
1743 Jean-Pierre Christin inverts the fixed points on Celsius' scale, to produce the scale used today.
2095: 2041: 1824: 1764: 1646: 474: 157: 819: 2090: 1954: 1693: 662:
Other electronic thermometers work by remote sensing: an infrared sensor responds to the radiation
458: 454: 207:
scale by setting the scale to the freezing and boiling points of water. By 1742 Swedish astronomer
181: 1989: 1615: 57: 1291:
Feverish illness in children: assessment and initial management in children younger than 5 years
2026: 1005: 961: 693: 689: 446: 1555: 925: 223:
was freezing. It was later reversed by Swedish botanist Carolus Linnaeus (1707–1778) in 1744.
2145: 2015: 1916: 898: 33:
A medical/clinical mercury thermometer showing the temperature of 37.7 °C (99.9 °F)
983: 2117: 2085: 2009: 1703: 1688: 1639: 593:) liquid crystals in a plastic strip that change color to indicate different temperatures. 231: 8: 1993: 1174: 903: 538: 508: 501: 497: 462: 247: 164:, he found, responded more quickly to temperature changes than the previously used water. 1130: 1113: 1979: 1754: 1718: 1497: 1470: 1399: 1374: 1229: 1143: 1054: 385:
in some countries or cultures, especially if used on patients older than young children
366: 329: 262: 250:, the "Thermometer of Lyon" built by the craftsman Pierre Casati that used this scale. 200: 193: 820:"History of the Thermometer: Timeline created by TheArctech in Science and Technology" 1926: 1911: 1906: 1813: 1683: 1678: 1580: 1502: 1404: 1330: 1221: 1213: 1178: 1135: 1094: 1090: 956: 933: 709: 530: 493: 258: 161: 139: 1233: 1147: 2003: 1944: 1850: 1673: 1492: 1482: 1422: 1394: 1386: 1322: 1205: 1170: 1125: 1086: 1020: 786: 278: 61: 1114:"The effect of hot beverages, cold beverages, and chewing gum on oral temperature" 881: 1865: 1860: 1819: 1785: 1252: 388: 129: 124:
The medical thermometer began as an instrument more appropriately called a water
28: 1949: 1934: 1805: 1390: 844:
Encyclopædia Britannica "Science & Technology: Daniel Gabriel Fahrenheit"
732: 371: 300: 208: 203:
created a clinical thermometer in 1665, to which he added an early form of the
1487: 154:(1610–1670), who created an enclosed thermometer that used alcohol circa 1654. 2134: 1723: 1708: 1698: 1217: 911: 773: 771: 769: 767: 765: 763: 761: 590: 413: 304: 266: 173: 64:
of a human or other animal. The tip of the thermometer is inserted into the
2036: 1969: 1964: 1870: 1845: 1800: 1584: 1506: 1408: 1334: 1139: 654:
reached, or the time may be specified in the manufacturer's documentation.
550: 405: 269:(1704–76). It was also utilized around the same time by Scottish physician 254: 1225: 1182: 1098: 989: 758: 526:
not very visible. Visibility is less of a problem with a coloured liquid.
1662: 1544: 894: 740: 685: 601: 375:
Different test prods (top: universal test prod, bottom: rectal test prod)
204: 185: 125: 1288:
National Collaborating Centre for Women's and Children's Health (2013).
1880: 1840: 1598:"An investigation into the accuracy of different types of thermometers" 1326: 1209: 857:"782 - Aerodrome reports and forecasts: A user's handbook to the codes" 284:
In 1868, German physician, pioneer psychiatrist, and medical professor
169: 1278:
Fundamentals of Nursing by Barbara Kozier et al., 7th edition, p. 495
1875: 1759: 1312: 1257: 907: 577: 557: 534: 296: 215:
temperature scale that was the reverse of the modern scale, in that
1795: 1749: 1739: 1013: 663: 450: 409: 146:
Two individuals switched from water to alcohol in the thermometer.
135: 109: 81: 1959: 1855: 1744: 553: 246:
represented boiling. On 19 May 1743 he published the design of a
212: 676:
reliable to replace the traditional glass/mercury thermometers"
1790: 1545:
Specification of typical inexpensive electronic ear thermometer
1050:"Dr. Theodor H. Benzinger, 94, Inventor of the Ear Thermometer" 874: 417: 393: 354: 289: 189: 89: 69: 1631: 1570: 967: 65: 814: 812: 810: 808: 806: 804: 560:
is actually used, resulting in a lower melting temperature.
227: 93: 1521:"How to use the Tempa DOT Single Use Clinical Thermometer" 299:, Germany, he immigrated to the U.S. in 1947 and became a 1044: 1042: 963:
MEMOIRE sur la dilatation du Mercure dans le Thermométre.
801: 563: 101: 199:
Prominent Dutch mathematician, astronomer and physicist
1076: 1002:
Memoires pour L'Histoire des Sciences et des Beaux Arts
785:. Oxford University Press. 1 April 2002. Archived from 1039: 453:. Temporal thermometers have been found to have a low 1354:. Oxford, England: Center for Evidence-Based Medicine 236:
Académie des sciences, belles-lettres et arts de Lyon
1375:"Thermometry and interpretation of body temperature" 722: 930:
The Facts on File Dictionary of Weather and Climate
703: 1195: 2132: 1163:Archives of Pediatrics & Adolescent Medicine 1347: 1027:. Academic Dictionaries and Encyclopedias. 2011 481: 152:Ferdinando II de' Medici, Grand Duke of Tuscany 1468: 1160: 1112:Newman, Bruce H.; Martin, Christin A. (2001). 1647: 1620:British and Foreign Medico-Chirurgical Review 838: 779:"A Brief History of the Clinical Thermometer" 314: 261:(1668–1738), as well as his notable students 1351:Accuracy of strip-like forehead thermometers 1111: 995: 468: 303:in 1955. He worked from 1947 to 1970 in the 242:represented the freezing point of water and 1573:Biomedical Instrumentation & Technology 966:Chaubert; Jean de Nully, Pissot, Duchesne, 1654: 1640: 949: 917: 581:Quick test based on thermo chromic colours 170:temperature scale which is named after him 27: 1496: 1486: 1398: 1129: 600: 576: 370: 1624:Vol.45, No.90, (April 1870), pp.429-441 1462: 1341: 880:Citation: Uppsala University (Sweden), 622:Resistance temperature detectors (RTDs) 2133: 1628:Vo.46, No.91, (July 1870), pp.144-156. 564:Phase-change (dot matrix) thermometers 226:Working independently of Celsius, the 219:was the boiling point of water, while 1999:Artificial intelligence in healthcare 1635: 1348:Brassey, Jon; Heneghan, Carl (2020). 1021:"Medical Dictionary: Ear Thermometer" 923: 541:and heat-sensitive liquid crystals). 238:, developed a similar scale in which 16:Device for measuring body temperature 1372: 1366: 1281: 1240: 1175:10.1001/archpedi.1996.02170260078013 973: 932:. Infobase Publishing. p. 246. 888: 679: 500:use a colored alcohol. Medically, a 253:The medical thermometer was used by 1469:Simpson, G.; Rodseth, R.N. (2019). 1248:"Hospitals: The Rectal Thermometer" 1131:10.1046/j.1537-2995.2001.41101241.x 13: 2074:reform debate in the United States 457:of around 60–70%, but a very high 436: 14: 2162: 2101:(Category Health care by country) 861:World Meteorological Organization 572: 324:, is discussed in the article on 178:territories and associated states 2113: 2112: 1091:10.1097/00003246-199609000-00012 725: 692:, the temperature upon waking. 558:alloy of gallium, indium and tin 486: 1661: 1591: 1564: 1556:"Takes Temperature in Seconds." 1549: 1538: 1513: 1441: 1427:Environmental Protection Agency 1415: 1306: 1272: 1189: 1154: 1105: 1070: 1010:Chaubert, París. pp. 2125-2128. 704:Smart and wearable thermometers 605:Electronic clinical thermometer 286:Carl Reinhold August Wunderlich 1810:Academic health science centre 1379:Biomedical Engineering Letters 1198:European Journal of Pediatrics 849: 381:water-based personal lubricant 1: 2047:Health information management 2032:health information technology 1770:Health information management 1610: 988:Imprimerie des Journalistes, 690:basal (base) body temperature 596: 520:Mercury-in-glass thermometers 326:normal human body temperature 309:Naval Medical Research Center 234:, permanent secretary of the 182:U.S. National Weather Service 2022:Translational bioinformatics 751: 617:Types of Digital Thermometer 482:Classification by technology 168:Fahrenheit also created the 7: 2052:Consumer health informatics 1315:Journal of Clinical Nursing 897:; Ronald A. Bailey (2005): 718: 670: 431: 279:Sir Thomas Clifford Allbutt 10: 2167: 1391:10.1007/s13534-019-00102-2 924:Smith, Jacqueline (2009). 900:Encyclopedia of Chemistry. 746:Liquid crystal thermometer 707: 648: 587:liquid crystal thermometer 544: 514: 364: 315:Classification by location 119: 2141:Medical testing equipment 2108: 2062: 2042:Public health informatics 1978: 1955:Electronic health records 1925: 1889: 1833: 1825:Supervised injection site 1778: 1765:Allied health professions 1732: 1669: 1618:, "Medical Thermometry", 1488:10.1186/s12871-019-0881-9 826:. Timetoast. January 1593 684:A basal thermometer is a 657: 589:contains heat-sensitive ( 475:plastic strip thermometer 469:Plastic strip thermometer 449:, and to a lesser degree 360: 348: 158:Daniel Gabriel Fahrenheit 42:Measures body temperature 38: 26: 21: 1561:, November 1954, p. 123. 926:"Appendix I: Chronology" 1990:Medical image computing 1940:Artificial intelligence 1294:. London, England: NICE 424:in medical facilities. 336: 311:in Bethesda, Maryland. 78:sub-lingual temperature 2027:Translational medicine 1423:"Mercury Thermometers" 1079:Critical Care Medicine 694:Basal body temperature 606: 582: 399: 376: 257:chemist and physician 2016:Computational biology 1917:Universal precautions 883:Linnaeus' thermometer 604: 580: 374: 56:is a device used for 2010:Behavior informatics 1373:Chen, Wenxi (2019). 498:alcohol thermometers 232:Jean-Pierre Christin 114:temporal temperature 106:tympanic temperature 86:axillary temperature 54:clinical thermometer 22:Clinical thermometer 1994:imaging informatics 1902:Cultural competence 1449:"Geratherm classic" 1429:. 21 September 2015 789:on 17 November 2014 509:thermal equilibrium 502:maximum thermometer 330:standard deviations 301:naturalized citizen 248:mercury thermometer 180:(all served by the 50:medical thermometer 2151:Italian inventions 1980:Health informatics 1755:Healthcare science 1475:BMC Anesthesiology 1327:10.1111/jocn.13117 1210:10.1007/BF02532519 1055:The New York Times 980:Journal helvétique 607: 583: 377: 367:Rectal thermometry 263:Gerard van Swieten 201:Christiaan Huygens 194:the Cayman Islands 98:rectal temperature 2126: 2125: 1890:Skills / training 1814:Teaching hospital 1603:, 1 October 2002. 1601:Nursing Times.net 1559:Popular Mechanics 1204:(12): 1005–1008. 1058:. 30 October 1999 957:Mercure de France 939:978-1-4381-0951-0 710:smart thermometer 688:used to take the 680:Basal thermometer 531:mercury poisoning 445:suffer issues of 259:Hermann Boerhaave 140:Santorio Santorio 128:, constructed by 46: 45: 2158: 2116: 2115: 2004:Neuroinformatics 1945:Connected health 1656: 1649: 1642: 1633: 1632: 1604: 1595: 1589: 1588: 1568: 1562: 1553: 1547: 1542: 1536: 1535: 1533: 1531: 1525: 1517: 1511: 1510: 1500: 1490: 1466: 1460: 1459: 1457: 1455: 1445: 1439: 1438: 1436: 1434: 1419: 1413: 1412: 1402: 1370: 1364: 1363: 1361: 1359: 1345: 1339: 1338: 1321:(7–8): 894–905. 1310: 1304: 1303: 1301: 1299: 1285: 1279: 1276: 1270: 1269: 1267: 1265: 1244: 1238: 1237: 1193: 1187: 1186: 1158: 1152: 1151: 1133: 1109: 1103: 1102: 1074: 1068: 1067: 1065: 1063: 1046: 1037: 1036: 1034: 1032: 1017: 1011: 999: 993: 977: 971: 970:. pp. 1609–1610. 953: 947: 946: 921: 915: 892: 886: 878: 872: 871: 869: 867: 853: 847: 842: 836: 835: 833: 831: 816: 799: 798: 796: 794: 775: 735: 730: 729: 728: 386: 307:division at the 150:The earliest is 62:body temperature 31: 19: 18: 2166: 2165: 2161: 2160: 2159: 2157: 2156: 2155: 2131: 2130: 2127: 2122: 2104: 2058: 2057: 2056: 1974: 1921: 1885: 1861:Overutilization 1829: 1820:Pharmacy school 1786:Assisted living 1774: 1728: 1665: 1660: 1613: 1608: 1607: 1596: 1592: 1569: 1565: 1554: 1550: 1543: 1539: 1529: 1527: 1523: 1519: 1518: 1514: 1467: 1463: 1453: 1451: 1447: 1446: 1442: 1432: 1430: 1421: 1420: 1416: 1371: 1367: 1357: 1355: 1346: 1342: 1311: 1307: 1297: 1295: 1286: 1282: 1277: 1273: 1263: 1261: 1246: 1245: 1241: 1194: 1190: 1159: 1155: 1110: 1106: 1075: 1071: 1061: 1059: 1048: 1047: 1040: 1030: 1028: 1019: 1018: 1014: 1000: 996: 978: 974: 954: 950: 940: 922: 918: 893: 889: 879: 875: 865: 863: 855: 854: 850: 843: 839: 829: 827: 818: 817: 802: 792: 790: 777: 776: 759: 754: 731: 726: 724: 721: 712: 706: 682: 673: 660: 651: 599: 575: 566: 547: 517: 489: 484: 471: 442:Temporal artery 439: 437:Temporal artery 434: 402: 384: 369: 363: 351: 339: 317: 130:Galileo Galilei 122: 34: 17: 12: 11: 5: 2164: 2154: 2153: 2148: 2143: 2124: 2123: 2121: 2120: 2109: 2106: 2105: 2103: 2098: 2093: 2088: 2083: 2081:United Kingdom 2078: 2077: 2076: 2066: 2064: 2060: 2059: 2055: 2054: 2049: 2044: 2039: 2034: 2029: 2024: 2019: 2013: 2007: 2001: 1996: 1986: 1985: 1984: 1982: 1976: 1975: 1973: 1972: 1967: 1962: 1957: 1952: 1950:Digital health 1947: 1942: 1937: 1935:3D bioprinting 1931: 1929: 1923: 1922: 1920: 1919: 1914: 1909: 1904: 1899: 1897:Bedside manner 1893: 1891: 1887: 1886: 1884: 1883: 1878: 1873: 1868: 1863: 1858: 1853: 1848: 1843: 1837: 1835: 1831: 1830: 1828: 1827: 1822: 1817: 1806:Medical school 1803: 1798: 1793: 1788: 1782: 1780: 1776: 1775: 1773: 1772: 1767: 1762: 1757: 1752: 1747: 1742: 1736: 1734: 1730: 1729: 1727: 1726: 1721: 1716: 1711: 1706: 1701: 1696: 1691: 1686: 1681: 1676: 1670: 1667: 1666: 1659: 1658: 1651: 1644: 1636: 1612: 1609: 1606: 1605: 1590: 1563: 1548: 1537: 1512: 1461: 1440: 1414: 1365: 1340: 1305: 1280: 1271: 1239: 1188: 1153: 1124:(10): 1241–3. 1104: 1069: 1038: 1025:enacademic.com 1012: 994: 992:. pp. 308-310. 972: 948: 938: 916: 887: 873: 848: 837: 800: 756: 755: 753: 750: 749: 748: 743: 737: 736: 733:Medical portal 720: 717: 708:Main article: 705: 702: 681: 678: 672: 669: 659: 656: 650: 647: 598: 595: 574: 573:Liquid crystal 571: 565: 562: 546: 543: 516: 513: 488: 485: 483: 480: 470: 467: 438: 435: 433: 430: 401: 398: 365:Main article: 362: 359: 350: 347: 338: 335: 316: 313: 271:George Martine 265:(1700–72) and 209:Anders Celsius 166: 165: 155: 121: 118: 44: 43: 40: 36: 35: 32: 24: 23: 15: 9: 6: 4: 3: 2: 2163: 2152: 2149: 2147: 2144: 2142: 2139: 2138: 2136: 2129: 2119: 2111: 2110: 2107: 2102: 2099: 2097: 2094: 2092: 2089: 2087: 2084: 2082: 2079: 2075: 2072: 2071: 2070: 2069:United States 2067: 2065: 2061: 2053: 2050: 2048: 2045: 2043: 2040: 2038: 2035: 2033: 2030: 2028: 2025: 2023: 2020: 2018:in healthcare 2017: 2014: 2012:in healthcare 2011: 2008: 2006:in healthcare 2005: 2002: 2000: 1997: 1995: 1991: 1988: 1987: 1983: 1981: 1977: 1971: 1968: 1966: 1963: 1961: 1958: 1956: 1953: 1951: 1948: 1946: 1943: 1941: 1938: 1936: 1933: 1932: 1930: 1928: 1924: 1918: 1915: 1913: 1910: 1908: 1905: 1903: 1900: 1898: 1895: 1894: 1892: 1888: 1882: 1879: 1877: 1874: 1872: 1869: 1867: 1864: 1862: 1859: 1857: 1854: 1852: 1849: 1847: 1844: 1842: 1839: 1838: 1836: 1832: 1826: 1823: 1821: 1818: 1815: 1811: 1807: 1804: 1802: 1799: 1797: 1794: 1792: 1789: 1787: 1784: 1783: 1781: 1777: 1771: 1768: 1766: 1763: 1761: 1758: 1756: 1753: 1751: 1748: 1746: 1743: 1741: 1738: 1737: 1735: 1731: 1725: 1722: 1720: 1717: 1715: 1712: 1710: 1707: 1705: 1702: 1700: 1697: 1695: 1692: 1690: 1687: 1685: 1682: 1680: 1677: 1675: 1672: 1671: 1668: 1664: 1657: 1652: 1650: 1645: 1643: 1638: 1637: 1634: 1630: 1629: 1625: 1621: 1617: 1616:Allbutt, T.C. 1602: 1599: 1594: 1586: 1582: 1579:(4): 259–65. 1578: 1574: 1567: 1560: 1557: 1552: 1546: 1541: 1522: 1516: 1508: 1504: 1499: 1494: 1489: 1484: 1480: 1476: 1472: 1465: 1450: 1444: 1428: 1424: 1418: 1410: 1406: 1401: 1396: 1392: 1388: 1384: 1380: 1376: 1369: 1353: 1352: 1344: 1336: 1332: 1328: 1324: 1320: 1316: 1309: 1293: 1292: 1284: 1275: 1259: 1255: 1254: 1253:Time Magazine 1249: 1243: 1235: 1231: 1227: 1223: 1219: 1215: 1211: 1207: 1203: 1199: 1192: 1184: 1180: 1176: 1172: 1168: 1164: 1157: 1149: 1145: 1141: 1137: 1132: 1127: 1123: 1119: 1115: 1108: 1100: 1096: 1092: 1088: 1085:(9): 1501–6. 1084: 1080: 1073: 1057: 1056: 1051: 1045: 1043: 1026: 1022: 1016: 1009: 1008: 1003: 998: 991: 987: 986: 981: 976: 969: 965: 964: 959: 958: 952: 945: 941: 935: 931: 927: 920: 913: 912:New York City 909: 905: 904:Facts On File 902: 901: 896: 891: 885: 884: 877: 862: 858: 852: 846: 841: 825: 824:Timetoast.com 821: 815: 813: 811: 809: 807: 805: 788: 784: 780: 774: 772: 770: 768: 766: 764: 762: 757: 747: 744: 742: 739: 738: 734: 723: 716: 711: 701: 697: 695: 691: 687: 677: 668: 665: 655: 646: 642: 641: 637: 633: 632: 631:Thermocouples 628: 624: 623: 619: 618: 614: 612: 603: 594: 592: 591:thermochromic 588: 579: 570: 561: 559: 555: 552: 542: 540: 536: 532: 527: 523: 521: 512: 510: 505: 503: 499: 495: 487:Liquid-filled 479: 476: 466: 464: 460: 456: 452: 448: 443: 429: 425: 421: 419: 415: 414:blood vessels 411: 407: 397: 395: 390: 389:Time Magazine 382: 373: 368: 358: 356: 346: 343: 334: 331: 327: 323: 322:clinical bias 312: 310: 306: 305:bioenergetics 302: 298: 293: 291: 287: 282: 280: 274: 272: 268: 267:Anton de Haen 264: 260: 256: 251: 249: 245: 241: 237: 233: 229: 224: 222: 218: 214: 210: 206: 202: 197: 195: 191: 187: 184:) as well as 183: 179: 175: 174:United States 171: 163: 159: 156: 153: 149: 148: 147: 144: 141: 137: 133: 131: 127: 117: 115: 111: 108:), or on the 107: 103: 99: 95: 91: 87: 83: 80:), under the 79: 75: 71: 67: 63: 59: 55: 51: 41: 37: 30: 25: 20: 2146:Thermometers 2128: 2037:Telemedicine 1970:Telemedicine 1965:Nanomedicine 1801:Nursing home 1619: 1614: 1600: 1593: 1576: 1572: 1566: 1558: 1551: 1540: 1528:. Retrieved 1515: 1478: 1474: 1464: 1452:. Retrieved 1443: 1431:. Retrieved 1426: 1417: 1382: 1378: 1368: 1356:. Retrieved 1350: 1343: 1318: 1314: 1308: 1296:. Retrieved 1290: 1283: 1274: 1262:. Retrieved 1260:8 April 1966 1251: 1242: 1201: 1197: 1191: 1166: 1162: 1156: 1121: 1117: 1107: 1082: 1078: 1072: 1060:. Retrieved 1053: 1029:. Retrieved 1024: 1015: 1006: 1001: 997: 984: 979: 975: 962: 955: 951: 943: 929: 919: 899: 890: 882: 876: 866:23 September 864:. Retrieved 860: 851: 840: 828:. Retrieved 823: 791:. Retrieved 787:the original 782: 713: 698: 683: 674: 661: 652: 643: 639: 638: 634: 630: 629: 625: 621: 620: 616: 615: 610: 608: 584: 567: 551:liquid metal 548: 528: 524: 518: 506: 490: 472: 440: 426: 422: 406:hypothalamus 403: 378: 353:The armpit ( 352: 344: 340: 321: 318: 294: 283: 275: 252: 243: 239: 235: 225: 220: 216: 211:created the 198: 167: 145: 134: 123: 113: 105: 100:), into the 97: 88:), into the 85: 77: 73: 53: 49: 47: 2096:New Zealand 1851:End-of-life 1733:Professions 1663:Health care 1454:14 December 1385:(1): 3–17. 1169:(1): 74–8. 1118:Transfusion 895:Don Rittner 741:Thermometer 686:thermometer 537:, coloured 459:specificity 455:sensitivity 387:; in 1966, 186:the Bahamas 126:thermoscope 2135:Categories 2063:By country 1927:Technology 1866:Palliative 1694:Philosophy 1684:Guidelines 1611:References 1530:23 October 1481:(1): 206. 1433:23 October 1358:23 October 1298:23 October 640:Thermistor 597:Electronic 428:detected. 230:physicist 138:physician 68:under the 2091:Australia 1912:Education 1907:Diagnosis 1760:Dentistry 1704:Providers 1679:Equipment 1674:Economics 1526:. BlueMed 1264:4 October 1258:Time Inc. 1218:1432-1076 990:Neuchâtel 914:. pp. 43. 908:Manhattan 752:Footnotes 535:galinstan 447:precision 297:Stuttgart 58:measuring 2118:Category 1796:Hospital 1779:Settings 1750:Pharmacy 1740:Medicine 1689:Industry 1585:11494651 1507:31706272 1409:30956877 1335:26994990 1234:21136002 1148:24681501 1140:11606822 1007:DE LYON. 1004:(1743): 982:(1743): 960:(1743): 719:See also 671:Accuracy 664:spectrum 539:alcohols 451:accuracy 432:Forehead 418:underarm 410:ear drum 290:underarm 228:Lyonnais 110:forehead 92:via the 1960:mHealth 1871:Primary 1856:Hospice 1846:Chronic 1745:Nursing 1714:Ranking 1498:6842509 1400:6431316 1226:8956933 1183:8542011 1099:8797622 1062:26 July 1031:26 July 830:16 July 793:26 July 649:Contact 611:digital 554:gallium 545:Gallium 515:Mercury 494:mercury 394:infants 213:Celsius 205:Celsius 162:Mercury 136:Italian 120:History 39:Purpose 2086:Canada 1791:Clinic 1724:System 1719:Reform 1709:Public 1699:Policy 1583:  1505:  1495:  1407:  1397:  1333:  1232:  1224:  1216:  1181:  1146:  1138:  1097:  936:  658:Remote 496:, but 361:Rectal 355:axilla 349:Armpit 192:, and 190:Belize 176:, its 90:rectum 82:armpit 70:tongue 1881:Total 1841:Acute 1524:(PDF) 1230:S2CID 1144:S2CID 985:LION. 968:Paris 255:Dutch 66:mouth 1992:and 1876:Self 1834:Care 1581:PMID 1532:2020 1503:PMID 1456:2021 1435:2020 1405:PMID 1360:2020 1331:PMID 1300:2020 1266:2022 1222:PMID 1214:ISSN 1179:PMID 1136:PMID 1095:PMID 1064:2016 1033:2016 934:ISBN 868:2009 832:2016 795:2016 337:Oral 94:anus 74:oral 60:the 1493:PMC 1483:doi 1395:PMC 1387:doi 1323:doi 1206:doi 1202:155 1171:doi 1167:150 1126:doi 1087:doi 783:QJM 463:ICU 412:'s 400:Ear 244:100 221:100 116:). 102:ear 76:or 52:or 2137:: 1812:, 1626:; 1622:, 1577:35 1575:. 1501:. 1491:. 1479:19 1477:. 1473:. 1425:. 1403:. 1393:. 1381:. 1377:. 1329:. 1319:25 1317:. 1256:. 1250:. 1228:. 1220:. 1212:. 1200:. 1177:. 1165:. 1142:. 1134:. 1122:41 1120:. 1116:. 1093:. 1083:24 1081:. 1052:. 1041:^ 1023:. 942:. 928:. 910:, 906:, 859:. 822:. 803:^ 781:. 760:^ 585:A 396:. 196:. 188:, 48:A 1816:) 1808:( 1655:e 1648:t 1641:v 1587:. 1534:. 1509:. 1485:: 1458:. 1437:. 1411:. 1389:: 1383:9 1362:. 1337:. 1325:: 1302:. 1268:. 1236:. 1208:: 1185:. 1173:: 1150:. 1128:: 1101:. 1089:: 1066:. 1035:. 870:. 834:. 797:. 240:0 217:0 112:( 104:( 96:( 84:( 72:(

Index


measuring
body temperature
mouth
tongue
armpit
rectum
anus
ear
forehead
thermoscope
Galileo Galilei
Italian
Santorio Santorio
Ferdinando II de' Medici, Grand Duke of Tuscany
Daniel Gabriel Fahrenheit
Mercury
temperature scale which is named after him
United States
territories and associated states
U.S. National Weather Service
the Bahamas
Belize
the Cayman Islands
Christiaan Huygens
Celsius
Anders Celsius
Celsius
Lyonnais
Jean-Pierre Christin

Text is available under the Creative Commons Attribution-ShareAlike License. Additional terms may apply.